View as PDF
* Main supply voltage fluctuations are not to exceed 10% of nominal supply voltage
** Other voltages available to meet International requirements
Handle down: 30 in (76 cm)
Handle up: 39.5 in (100 cm)
Length: 22 in (56 cm)
Width: 12.5 in (32 cm)
Empty: 105 lbs (48 kg)
With reservoir filled: 116 lbs (53 kg)
115 VAC, 60 Hz, 11.0 Amp (Nominal)**
Current leakage: < 300uA
Circuit breaker: 12.0 Amp
5.5 x 8.75 in (14 x 22 cm) display
360º rotation
Two available, YSI 400 Series compatible
Serial RS232
5 liters
Circulating Fluid
Sterile or distilled water
Audible tones and displayed text messages
15 ft (4.6 m)
10 ft (3.0 m)
Operating: 10 to 27ºC, 50 to 80ºF
Storage: -30 to 50ºC, -20 to 120ºF
Operating: 5-70%
Storage: 5-95%, non-condensing
Automatic, Manual, Purge, Stop
750 W
Distilled or Sterile Water
5 liters
Two available, YSI 400 Series compatible
YSI 400 Series compatible
Temp Input 1 - Control/Monitor
Temp Input 2 - Monitor
10°C to 44°C, 50°F to 111.2°F
in 0.1°C/ºF
±0.4ºC 10 to 32ºC
±0.2ºC 32 to 38ºC
±0.4ºC 38 to 44ºC
includes ± 0.1ºC external probe
32°C to 38.5°C
89.6°F to 101.3°F
in 0.1°C/F increments
3°C to 45°C
37.4°F to 113.0°F
in 1°C/F
4°C to 42°C
39.2°F to 107.6°F
in 0.1°C/F increments