Arctic Sun > Performance Specifications

Control Modes
Automatic, Manual, Purge, Stop

Heater Capability
750 W

Circulating Fluid
Distilled or Sterile Water

Reservoir Capacity
5 liters

Water flow rate (total)
0.5 - 8.0 liter/min

Patient Probe Type
YSI 400 Series compatible

Patient Temperature Inputs (2)
Temp Input 1 - Control/Monitor
Temp Input 2 - Monitor

Patient Temperature Display Range
10°C to 44°C, 50°F to 111.2°F
in 0.1°C/ºF

 

Patient Temperature Measurement Accuracy
±0.4ºC 10 to 32ºC
±0.2ºC 32 to 38ºC
±0.4ºC 38 to 44ºC
includes ± 0.1ºC external probe

Patient Temperature Control Range - Automatic Mode
32°C to 38.5°C
89.6°F to 101.3°F
in 0.1°C/F increments

Water Temperature Display Range
3°C to 45°C
37.4°F to 113.0°F
in 1°C/F

Water Temperature Control Range - Manual Mode
4°C to 42°C
39.2°F to 107.6°F
in 0.1°C/F increments